Dose Calibrators

  • Introduction
  • Standard quality control tests for radionuclide calibrators
  • Frequency of tests and Acceptable tolerances

There is a set hierarchy in terms of accuracy of dose calibrators. The Primary standard is held and maintained at the National Physics Laboratory (NPL). Secondary standards are those which have been directly cross calibrated with the primary. Tertiary standards (or field instruments) are cross calibrated with secondary standards. It is the tertiary standards that are generally held in nuclear medicine departments.

 

The frequency and tolerances outlined below are taken from the IPSM (now the IPEM) report no.65 Quality Standards in Nuclear Medicine. It should be however be noted that new European guidelines are now being drafted.

Table 1: Frequency of tests
  Daily Monthly Yearly Acceptance testing
High voltage x x x x
Zero adjust x x x x
Background x x x x
Accuracy x x x x
Relative responses x x x x
Precision   x x x
Linearity     x x
Electrical safety     x x
Subsidiary calibrations       x
Leakage radiation       x

 

Table 2: Acceptable tolerances
  Parameter Reference Instrument Field Instrument
  Precision ±0.5% ±1%
  Reproducibility ±1%

±2% sealed chamber

±3% open chamber
  Linearity ±1% corrected ±5% over range used
Accuracy Agreement with secondary standard

±2% high energy

±5% low energy
 
Agreement with tertiary standard  

±5% high energy

±10% low energy

*Low energy refers to Gamma or x-ray emitters where more than 30% of the emissions are below 100keV.